Rayner Sulcoflex® Pseudophakic Supplementary IOL
With advanced IOL design and modern surgical technique, an exact refractive result following cataract surgery with implantation of an IOL is a reasonable expectation.

Rayner Sulcoflex® Pseudophakic Supplementary IOLs are designed to correct any post-surgical or residual ametropia, without the trauma, or increased surgical risk, of an IOL exchange.

Moreover, unlike with conventional "piggy-back" IOL implantations, the unique design of Rayner Sulcoflex® Pseudophakic Supplementary IOLs ensures that the potential for contact between the two implants is minimised. This reduces the likelihood of induced refractive error and optical aberrations due to optic surface distortion at the area of contact between the IOLs.

Rayner Sulcoflex® Pseudophakic Supplementary IOLs are manufactured from Rayner’s proprietary hydrophilic acrylic co-polymer, Rayacryl®, which is noted for its exceptional uveal biocompatibility – a factor so important in those IOLs specifically intended for ciliary sulcus fixation. Rayacryl® is also noted for its superb optical clarity, with no vacuoles or glistenings. These factors, coupled with a remarkably low silicone oil adherence and excellent handling characteristics, means that Rayacryl® represents the ideal in IOL biomaterials.

The large optic diameter of Rayner Sulcoflex® Pseudophakic Supplementary IOLs ensures quality of vision with reduced risk of optic-iris capture and the round optic edge contributes to low rates of dysphotopsia.

The large diameter “undulating” haptics ensures IOL centration and rotational stability and the haptic angulation ensures uveal clearance, thereby avoiding pigment dispersion, and further reducing the risk of optic-iris capture.

The soft, round-edged haptic design ensures minimum tissue contact reaction.

Rayner Sulcoflex® Pseudophakic Supplementary IOLs
are specifically indicated for:

The correction of Post-surgical ametropia
Enhancement of the refractive result after RLE or PRELEX.
Enhancement of near / far vision
Correction of any residual corneal astigmatism
Correction of refractive changes following paediatric implantation.

Physical parameters
Optic Body Diameter 6.5mm
Overall Length 13.5mm
Haptic angulation 10°
Optic configuration Anterior Convex, Posterior Concave


Power Availability
Sulcoflex® Aspheric (653L)
Standard range
-5.0D to +5.0D in 0.5D increments
Premium range*
-10.0D to -5.5D in 0.5D increments
+5.5D to +10.0D in 0.5D increments

Sulcoflex® Multifocal (653F)
-3.0D to +3.0D in 0.5D increments with +3.5D add.

Sulcoflex® Toric (653T)
Standard range
SE –3.0D to +3.0D in 0.5D steps, cylinder +1D to +3D in
1.0D steps.

Cylinder 1.0D 2.0D 3.0D
Minimum sphere -3.5D -4.0D -4.5D
Maximum sphere +2.5D +2.0D +1.5D

Premium range*
SE –6.0D to +6.0D in 0.5D steps, cylinder +1D to +6D in
0.5D steps.

Cylinder 1.0D 1.5D 2.0D 2.5D 3.0D 3.5D 4.0D 4.5D 5.0D 5.5D 6.0D
Minimum sphere -6.5D -6.5D -7.0D -7.0D -7.5D -7.5D -8.0D -8.0D -8.5D -8.5D -9.0D
Maximum sphere +5.5D +5.0D +5.0D +4.5D +4.5D +4.0D +4.0D +3.5D +3.5D +3.0D +3.0D

*Manufactured specifically to order.


Presentation
Rayner Sulcoflex® Pseudophakic Supplementary IOLs are supplied in 0.9% saline solution in a pouched blister pack, sterilised by moist heat and presented in a convenient, ready to use Lens Injection System Pack containing:
One Sulcoflex® Pseudophakic Supplementary IOL
One Rayner Single Use Soft-Tipped Injector

Sulcoflex® Pseudophakic Supplementary IOLs are exclusively designed in collaboration with Professor Michael Amon (Vienna, Austria). Sulcoflex® patent pending.

Click here to download brochure.

Click here to download Eurotimes article - November 2007.

Click here to download article by Professor MICHAEL AMON
(CATARACT & REFRACTIVE SURGERY TODAY EUROPE - MARCH 2009)

Click here to download Sucoflex Loading Instructions

If you would like to register to use the Raytrace™ software program for personalised toric IOL calculations, please contact your local Designs for Vision representative or register via the following link
http://www.rayner.com/raytrace/register

More information on the ordering system, see this brochure (Warning: 1.4MB .pdf).