Scotopic Sensitivity Tester

SST-1
Whole-field scotopic sensitivity (threshold) testing provides clinically beneficial information in a wide variety of ophthalmic conditionals. Dark adaptometry is useful in the diagnosis and management of retinal degeneration, senile miosis, high myopia vitamin A deficiency, and other night blinding conditions, while whole-field scotopic sensitivity has been shown to be useful in the early detection of glaucoma.

SST-1 Features:
  • Dark Adaptometry Mode - Measures conventional dark adaptometry curve of final dark-adapted threshold.
  • Whole Field Scotopic Sensitivity Mode - measures dark-adapted sensitivity to 1 Hz sinusoidal flicker.
  • Ganzfeld Stimulus - test entire retina
  • Fast - Whole field scoptopic sensitivity or final dark-adapted threshold can be determined in 1-2 minutes.
  • Inexpensive - Increased functionality at a fraction of the price.
  • Simple - No need for the patient to maintain accurate fixation. No requirement for refractive correction.
  • Results are relatively unaffected by mild cataract.
  • Small and portable - Lightweight (4lbs) and compact. Fits easily into briefcase sized travelling case (optional).
  • Easy to Use - All necessary controls are on the hand held Ganzfeld Stimulator Unit
The SST-1 is a flexible solution, using innovative high technology, to the problems of determining dark adaptation curves and whole field scotopic sensitivity.


More Information:

Configuration :
Base: 4" H x 8" W x 9" D (Approximately 10 cm x 20 cm x 22 cm); 3 lbs.
Stimulator: 9" L x 3" W x 3" D (Approximately 22 cm x 7 cm x 7 cm);
                  less than 1 lb.
Power: 115/220 V 50/60 Hz, 20 Watts

Dark Adaptometry :
Pre-adapting field of 1,000 cd "m-2 for 60 seconds (timed). May be bypassed for measurement of final dark-adapted threshold.
Timer resolution is 0.1 minutes. Will measure time to several hours.
Stimulus is ½ second flash.
Intensity is adjustable over a 3 log unit range in 0.1 log unit steps.

Whole-Field Scotopic Sensitivity :
Implements modified test of Glovinksy, et al.1
No need for steady fixation or refractive correction. Results are relatively unaffected by mild cataracts.
Stimulus is 4 seconds or 1 Hz sine wave flicker.
Intensity is adjustable over a 3 log unit range in 0.1 log unit steps.

Reference :
Glovinsky Y, Quigley HA, Drum B, Bisset RA, Jampel HD. A whole-field scotopic retinal sensitivity test for the detection of early glaucoma damage. Archives of Ophthalmology 1992; 110:486-490

Example :
Dark Adaptometry Curve for Normal and
Congenital Stationary Night Blindness

More Description:
Whole-field scotopic sensitivity (threshold) testing provides clinically beneficial information in a wide variety of ophthalmic conditions. Dark adaptometry is useful in the diagnosis and management of retinal degenerations, senile miosis, high myopia, vitamin A deficiency, and other night blinding conditions. It has also been shown to be useful in the early detection of glaucoma1. The SST-1 is a flexible solution to the problems of determining dark adaptation curves and whole-field scotopic sensitivity. The SST-1 is versatile and easy to use. The controls are conveniently located on the hand-held stimulator unit. The base unit displays the stimulus and intensity, and the time elapsed when in dark adaptometry mode. The SST-1 is lightweight, portable, and easily stored when not in use. Minimal operator training is required to use the SST-1 in either dark adaptometry or whole-field scotopic sensitivity mode.

In addition to functioning as a dark adaptometer, the SST also implements the whole-field scotopic sensitivity test developed by clinicians at the prestigious Wilmer Eye Institute. This test measures the dark-adapted threshold to a 1 Hz flickering light presented in the ganzfeld. Glovinsky et al.1, used a laboratory version of the SST-1 to show that glaucoma suspects with an abnormal whole-field scotopic sensitivity are more likely to have other signs of early optic nerve injury, including a higher proportion of borderline visual field defects, defects of the optic nerve fiber layer, or glaucomatous fellow eyes. Unlike conventional perimetry, the accuracy of the SST-1 is not affected by poor refractive error correction or erratic fixation. The ability of the SST-1 to record meaningful data is also relatively insensitive to mild cataracts because the entire visual field is stimulated.

In the whole field scotopic sensitivity test, one eye is dark adapted for 30 minutes. The luminance threshold to a 1 Hz sine wave flicker is determined by adjusting the brightness of the stimulus until it is just visible to the patient. The time for the entire procedure (excluding dark-adapt time) is 1 to 2 minutes.

 
     